アブストラクト(18巻2号:The Bulletin of Kanagawa Dental College)

The Bulletin of Kanagawa Dental College

English

Title : A Study of Recording Accuracy in Linear Measurements Displayed by Sirognathograph
Subtitle : SHORT COMMUNICATION
Authors : Yukari Endo, Hideo Aoki, Tadahiro Fujita, Masaaki Yamamura, Noriko Takeuchi, Katsuhiko Kimoto, Katsushi Tamaki, Yoichi Nagashima
Authors(kana) :
Organization : Department of Crown and Bridge Prosthodontics, Kanagawa Dental College
Journal : The Bulletin of Kanagawa Dental College
Volume : 18
Number : 2
Page : 103-105
Year/Month : 1990 / 9
Article : Report
Publisher : Kanagawa Odontological Society
Abstract : [Abstract] The recording accuracy of Sirognathograph (Siemens, West Germany) was conducted by means of an experimental apparatus. Diagrams of mandibular border movements in three dimensional directions were performed. The results showed the distortion was not recognized within 5mm of primary condition in a subject, but slightly between 5mm and 10mm. In a distance of more than 15mm, a considerable distortion was increased until at the maximum opening position.
Practice : Dentistry
Keywords : Recording Accuracy, Mandibular Border Movements, Three Dimensional Evaluation, Sirognathograph