Title : | Basic Study of Filters for One-shot Dual Energy Subtraction Using Iodinated Contrast Medium |
---|---|
Subtitle : | SHORT COMMUNICATION |
Authors : | Hiromi Wakao, Satoshi Kobayashi, Takahiko Matsuki, Yasuo Shirase, Joji Yoneda, Michiharu Shimamoto, Haruyoshi Kuriyama, Isamu Kashima |
Authors(kana) : | |
Organization : | Department of Oral Radiology, Kanagawa Dental College |
Journal : | The Bulletin of Kanagawa Dental College |
Volume : | 24 |
Number : | 1 |
Page : | 25-27 |
Year/Month : | 1996 / 3 |
Article : | Report |
Publisher : | Kanagawa Odontological Society |
Abstract : | [Abstract] Fundamental studies of one-shot dual energy subtraction were performed to obtain images of contrast medium alone by eliminating bone images. However, subtraction in actual images of the body requires elimination of images of not only bone, but also of soft tissue, which continuously alters in thickness. In the present study, we examined changes in X-ray transmission to contrast medium, bone, and soft tissue as tube voltage and a filter for energy separation were changed. It was revealed that a contrast medium, for which conventional iodine was used, was sufficiently separated not only from bone, but from soft tissue as well, when a tube voltage of 50 kVp and a 0.8 mm copper filter were used. |
Practice : | Dentistry |
Keywords : | One-shot dual energy subtraction, Filter, K-edge |